Experimental set up for the study of plasma parameters in seeded arc plasma using Langmuir Probes inside low-cost vacuum chamber

Authors

  • Vijay Kumar Jha Central Department of Physics, Kirtipur, Tribhuvan University, Kathmandu
  • Lekha Nath Mishra Patan Multiple Campus, Patandhoka, Lalitpur, Tribhuvan University, Kathmandu
  • Bijoyendra Narayan Dr. S. K. Sinha Women’s College, Motihari, B. R. A. Bihar University, Muzaffarpur

DOI:

https://doi.org/10.3126/bibechana.v16i0.20839

Keywords:

Arc plasma, vacuum chamber, plasma parameters, operational constraints, Langmuir Probe, floating Potential, electron density.

Abstract

In this piece of study, the design and development of a low cost vacuum chamber with different ports are mentioned. I-V characteristic of Langmuir Probe is obtained using the primary data of the experimental set up for ‘single probe method’ in nitrogen seeded arc plasma at atmospheric pressure The floating potential was found to be 36V and the electron density to be 1.24 ´ 104 K. Variation in electron density of the arc plasma with probe potential is also studied at low pressure range from 10 Nm-2 (0.10 mbar) to 20 Nm-2.

BIBECHANA 16 (2019) 1-6

Downloads

Download data is not yet available.
Abstract
2465
PDF
635

Downloads

Published

2018-11-22

How to Cite

Jha, V. K., Mishra, L. N., & Narayan, B. (2018). Experimental set up for the study of plasma parameters in seeded arc plasma using Langmuir Probes inside low-cost vacuum chamber. BIBECHANA, 16, 1–6. https://doi.org/10.3126/bibechana.v16i0.20839

Issue

Section

Research Articles