Interference of Bilirubin in Creatinine Value Measurement by Jaffe Kinetic Method
DOI:
https://doi.org/10.3126/acclm.v1i1.12311Keywords:
Creatinine, True creatinine, Bilirubin total, Bilirubin direct (T/D), out valueAbstract
BACKGROUND: Creatinine measurement in icteric sample is a major but unresolved problem. Bilirubin causes negative interference in creatinine value measurement using general techniques. The objective of this study was to find differences in creatinine value by Jaffe Kinetic method pre-incubation and without pre-incubation with NaOH.
METHODS: This was a cross sectional, descriptive study carried out in 71 samples with different level of bilirubin concentration. We took blood samples of 71 different patients, 48 males and 23 females, from two different hospitals of Kathmandu village. Both creatinine and bilirubin concentration in serum samples were measured by using Staxfax 3300 semi auto analyzer in the hospital. In the laboratory creatinine value was measured by kinetic method and bilirubin measured by Jendrassik/ Grof method using commercial kits. Statistical analysis of quantitative data was done by using SPSS version 16.0.
RESULTS: The results shows differences in creatinine values with respect to methods and extent of bilirubin concentration. It was found that the creatinine obtained by pre-incubation with NaOH has greater value than without pre-incubation (i.e. by direct estimation using working reagent). It was also shown that the high bilirubin cause the interference in greater extent. The significant interference was seen in the sample with bilirubin concentration greater than 20 mg/dl i.e. creatinine value after treatment with NaOH prior to dispensing picric acid is significantly increases, P<0.01 at 99% confidence level.
CONCLUSION: This shows that the bilirubin has negative interference in creatinine value measurement by ordinary laboratory practices and interference increases with higher concentration of bilirubin in blood sample.
DOI: http://dx.doi.org/10.3126/acclm.v1i1.12311
Ann. Clin. Chem. & Lab. Med. 1(1) 2015: 25-28
Downloads
Downloads
Published
How to Cite
Issue
Section
License
Authors who publish with this journal agree to the following terms:
- The author transfers copyright to the Nepalese Association for Clinical Chemistry.
- The journal publishes the work under a Creative Commons Attribution License that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal and under the same share-alike license used here.
- Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
- Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).